Power supply (supply voltage suffix code -2), Isolation – Yokogawa Button Operated MV2000 User Manual
Page 277
13-15
M-4660
Specifications
1
2
3
4
5
6
7
8
9
10
11
12
13
App
Index
Power Supply (supply voltage suffix code -1)
Item
Specifications
Rated supply voltage
100 to 240 VAC
Supply voltage range
90 to 132, 180 to 264 VAC
Rated power supply frequency 50 Hz or 60 Hz
Power consumption
RD-MV1000/RD-
MV2000
Supply voltage
LCD backlight off
Normal
Maximum
RD-MV1000
100 VAC
15 VA
30 VA
45 VA
240 VAC
25 VA
40 VA
60 VA
RD-MV2000
100 VAC
28 VA
40 VA
65 VA
240 VAC
38 VA
54 VA
90 VA
Allowable interruption time
Less than 1 cycle of the power supply frequency
Power Supply (supply voltage suffix code -2)
Item
Specifications
Rated supply voltage
12 VDC/24 VDC
Supply voltage range
10.0 to 28.8 VDC
Power consumption
RD-MV1000/RD-
MV2000
Supply voltage
LCD backlight off
Normal
Maximum
RD-MV1000
12 VDC
7 VA
14 VA
24 VA
24 VDC
7 VA
14 VA
23 VA
RD-MV2000
12 VDC
9 VA
18 VA
35 VA
24 VDC
9 VA
18 VA
33 VA
Miscellaneous
A/D converter integration time: Fixed to 20 ms (50 Hz) if set to Auto
When using the AC adapter
Rated supply voltage
100 to 240 VAC
Supply voltage range
90 to 264 VAC
Rated power supply frequency 50 Hz or 60 Hz
Maximum input power consumption
RD-MV1000: 60 VA, RD-MV2000: 90 VA
Rated power supply frequency 48 to 62 Hz
Power consumption
RD-MV1000/RD-
MV2000
Supply voltage
LCD backlight off
Normal
Maximum
RD-MV1000
100 VAC
15 VA
30 VA
45 VA
240 VAC
25 VA
40 VA
60 VA
RD-MV2000
100 VAC
28 VA
40 VA
65 VA
240 VAC
38 VA
54 VA
90 VA
Dielectric strength
1500 VAC (50/60 Hz) for one minute between the AC adapter power supply terminal and
earth
Miscellaneous
A/D converter integration time: Fixed to 20 ms (50 Hz) if set to Auto
Isolation
Item
Specifications
Insulation resistance
Between the Ethernet, RS-422/485, and insulation terminals and earth: 20 MΩ or greater at
500 VDC
Dielectric strength
Between the power terminal and earth: 2300 VAC at 50/60 Hz for one minute
Between the contact output terminal and earth: 1600 VAC at 50/60 Hz for one minute
Between the signal input terminal and earth: 1500 VAC at 50/60 Hz for one minute
Between signal input terminals: 1000 VAC (50/60 Hz) for one minute (excluding the “b”
input terminal of RD-MV1006, RD-MV1012, RD-MV1024,
RD-MV2010, RD-MV2020, RD-MV2030, RD-MV2040, RD-
MV2048)
Between the remote input terminal and earth: 1000 VDC for one minute
Between the pulse input terminal and earth: 1000 VDC for one minute
Ground
Grounding resistance: 100 Ω or less
13.6 General Specifications