beautypg.com

App index standard performance – Yokogawa Button Operated MV2000 User Manual

Page 279

background image

13-17

M-4660

Specifications

1

2

3

4

5

6

7

8

9

10

11

12

13

App

Index

Standard Performance

Item

Specifications

Measurement/display accuracy

Standard operating conditions:

Temperature:

23 ± 2°C

Humidity:

55% ± 10%RH

Power supply voltage:

90 to 132 or 180 to 250 VAC

Power supply frequency:

50/60 Hz ± 1%

Warm-up time:

At least 30 minutes.

Other ambient conditions such as vibration should not adversely affect the operation.

Input Type

20 mV
60 mV
200 mV
2 V
1-5 V
6 V
20 V
50 V
R

S

B

K

E
J
T
N
W
L
U
WRe

Pt100
JPt100

A/D integration time: 16.7 ms or more
(0.05% of rdg + 12 digits)

(0.05% of rdg + 3 digits)

(0.05% of rdg + 12 digits)

(0.05% of rdg + 3 digits)

(0.15% of rdg + 1°C)
R, S

0 to 100°C: 3.7°C,

100 to 300°C: 1.5°C

B

400 to 600°C: 2°C,

Accuracy not guaranteed for

values less than 400°C

(0.15% of rdg + 0.7°C)
–200 to –100°C: (0.15% of rdg + 1°C)
(0.15% of rdg + 0.5°C)
–200 to –100°C: (0.15% of rdg + 0.7°C)

(0.15% of rdg + 0.7°C)
(0.15% of rdg + 1°C)
(0.15% of rdg + 0.5°C)
–200 to –100°C: (0.15% of rdg + 0.7°C)
(0.2% of rdg + 2.5°C)
0 to 200°C: 4.0°C

(0.15% of rdg + 0.3°C)

A/D integration time: 1.67 ms
(0.1% of rdg + 40 digits)

(0.1% of rdg + 15 digits)

(0.1% of rdg + 40 digits)

(0.1% of rdg + 15 digits)

(0.2% of rdg + 4°C)
R, S

0 to 100°C: 10°C,

100 to 300°C: 5°C

B

400 to 600°C: 7°C,

Accuracy not guaranteed for

values less than 400°C

(0.2% of rdg + 3.5°C)
–200 to –100°C: (0.15% of rdg + 6°C)
(0.2% of rdg + 2.5°C)
–200 to –100°C: (0.2% of rdg + 5°C)

(0.3% of rdg + 3.5°C)
(0.3% of rdg + 7°C)
(0.2% of rdg + 2.5°C)
–200 to –100°C: (0.2% of rdg + 5°C)
(0.3% of rdg + 10°C)
0 to 200°C: 18.0°C

(0.3% of rdg + 1.5°C)

DC voltage

Thermocouple

• Not including the

accuracy of reference

junction compensation

• With burnout detection

function OFF

RTD

Max. Resolution
of Digital Display
1 µV
10 µV
10 µV
100 µV
1 mV
1 mV
1 mV
10 mV

0.1°C

Measurement Accuracy (Digital Display)

Range

Voltage

Contact

Threshold level (Vth=2.4 V) accuracy ± 0.1 V
1 kΩ or less: 1 (ON), 100 kΩ or more: 0 (OFF) (parallel capacitance of 0.01 µF or less)

DI

13.6 General Specifications

This manual is related to the following products: