App index standard performance – Yokogawa Button Operated MV2000 User Manual
Page 279
13-17
M-4660
Specifications
1
2
3
4
5
6
7
8
9
10
11
12
13
App
Index
Standard Performance
Item
Specifications
Measurement/display accuracy
Standard operating conditions:
Temperature:
23 ± 2°C
Humidity:
55% ± 10%RH
Power supply voltage:
90 to 132 or 180 to 250 VAC
Power supply frequency:
50/60 Hz ± 1%
Warm-up time:
At least 30 minutes.
Other ambient conditions such as vibration should not adversely affect the operation.
Input Type
20 mV
60 mV
200 mV
2 V
1-5 V
6 V
20 V
50 V
R
S
B
K
E
J
T
N
W
L
U
WRe
Pt100
JPt100
A/D integration time: 16.7 ms or more
(0.05% of rdg + 12 digits)
(0.05% of rdg + 3 digits)
(0.05% of rdg + 12 digits)
(0.05% of rdg + 3 digits)
(0.15% of rdg + 1°C)
R, S
0 to 100°C: 3.7°C,
100 to 300°C: 1.5°C
B
400 to 600°C: 2°C,
Accuracy not guaranteed for
values less than 400°C
(0.15% of rdg + 0.7°C)
–200 to –100°C: (0.15% of rdg + 1°C)
(0.15% of rdg + 0.5°C)
–200 to –100°C: (0.15% of rdg + 0.7°C)
(0.15% of rdg + 0.7°C)
(0.15% of rdg + 1°C)
(0.15% of rdg + 0.5°C)
–200 to –100°C: (0.15% of rdg + 0.7°C)
(0.2% of rdg + 2.5°C)
0 to 200°C: 4.0°C
(0.15% of rdg + 0.3°C)
A/D integration time: 1.67 ms
(0.1% of rdg + 40 digits)
(0.1% of rdg + 15 digits)
(0.1% of rdg + 40 digits)
(0.1% of rdg + 15 digits)
(0.2% of rdg + 4°C)
R, S
0 to 100°C: 10°C,
100 to 300°C: 5°C
B
400 to 600°C: 7°C,
Accuracy not guaranteed for
values less than 400°C
(0.2% of rdg + 3.5°C)
–200 to –100°C: (0.15% of rdg + 6°C)
(0.2% of rdg + 2.5°C)
–200 to –100°C: (0.2% of rdg + 5°C)
(0.3% of rdg + 3.5°C)
(0.3% of rdg + 7°C)
(0.2% of rdg + 2.5°C)
–200 to –100°C: (0.2% of rdg + 5°C)
(0.3% of rdg + 10°C)
0 to 200°C: 18.0°C
(0.3% of rdg + 1.5°C)
DC voltage
Thermocouple
• Not including the
accuracy of reference
junction compensation
• With burnout detection
function OFF
RTD
Max. Resolution
of Digital Display
1 µV
10 µV
10 µV
100 µV
1 mV
1 mV
1 mV
10 mV
0.1°C
Measurement Accuracy (Digital Display)
Range
Voltage
Contact
Threshold level (Vth=2.4 V) accuracy ± 0.1 V
1 kΩ or less: 1 (ON), 100 kΩ or more: 0 (OFF) (parallel capacitance of 0.01 µF or less)
DI
13.6 General Specifications