Execution errors – Yokogawa Button Operated MV2000 User Manual
Page 243
12-4
M-4660
• Execution Errors
Code
Message
Explanation/Countermeasures/Ref. section
150
This action is not possible because sampling is in
progress.
Stop memory sampling and then execute.
See section 4.4
151
This action is not possible during sampling or calculating.
Stop memory sampling or computation first.
See sections 4.4 and 10.4
152
This action is not possible because saving is in progress.
Wait until the saving is complete.
153
This action is not possible because formatting is in
progress.
Wait until the formatting is complete.
154
Message not accepted because message limit was
reached.
The limit is 50 messages.
See section 6.4
155
The message is not written while sampling is stopped.
Start memory sampling and then execute.
See section 4.4
156
There are no channels to be saved to the memory.
Set the channels to be saved.
See sections 4.1 and 10.1
157
This function is not possible at this time.
If the data type to be recorded is set to E+D, you cannot
change the trend interval by using the T/DIV key.
See section 5.2
158
Exceeds time deviation setting.
When synchronizing the clock through remote control.
See section 1.7
159
It is outside the postscript message write-in range.
Add message can be written to the past section of the
data being memory sampled.
See section 6.4
12.1 List of Messages