background image

4-20

M-4660

2.

Select the setup file to be read by using the arrow keys and DISP/ENTER.

* The setup data file is in the root directory (

/

).

To cancel the operation, press ESC.

3.

Press the DISP/ENTER key.

The setup data is loaded.

Explanation

• Setup Data Files

• The setup data file extension is .PDL.
• The maximum setup data file size is about 200 KB.
• The following settings are also saved.
• The current monitor display condition.
• The Jump default display setting.
• Favorite display key data.

• Loading Setup Data

• The monitor display condition, Jump default display setting, and favorite display key

settings are also loaded.

• If the loaded setup data is not applied, check the error log (see section 5.10).
• While setup data is being loaded, key operations, communication operations, and

remote control operations cannot be executed.