beautypg.com

26 probe settings, Setting auto calibration – Teledyne LeCroy Summit T24 PCIe Multi-lane Protocol Analyzer User Manual User Manual

Page 103

background image

 

Summit T24 PCI Express Multi‐Lane Protocol Analyzer User Manual

91

Probe Settings

 Teledyne LeCroy

5.26 Probe

Settings

The Summit T24 analyzer can be optimized for specific DUT signal characteristics. The 
auto calibration feature in the Probe Settings allows for better signal recovery. There are 
16, 0 db to 15 db, internal equalization modes in the analyzer hardware that boost the 
high frequency content of the captured signal. This equalization does not impact the 
signal to the DUTs, but helps in ‘opening the eye’ of the captured signal, so that the 
internal logic can capture PCIe frames. The system is designed to allow you to manually 
set the equalization to one of the sixteen modes for the CTLE values and one of the eight 
modes for the DFE values or choose to run an automated algorithm, internal to the 
analyzer. This automated calibration sets the optimized values for each of the PCIe 
captured lanes, provided that the signal at the analyzer input is stable.

Two lane settings can be configured, CTLE and DFE (Decision Feedback Equalization). CTLE 
is two dimensional and is set based on the least number of errors returned. DFE is three 
dimensional and is set based on the least number of errors returned by both CTLE and 
DFE.

Select Recording Options > Probe Settings from the Setup menu to display Probe 
Settings.

Setting Auto Calibration

The Probe Settings tab in the Recording Options dialog displays the equalizer settings. To 
set all lanes upstream and downstream check the Set All Lanes boxClick on Auto 
Configure 
to display the Calibrate dialog (see 

Figure 5.50 on page 93

).