Model 9110TH NOx Analyzer

Operating

Instructions

Teledyne Analytical Instruments

60

SAMPLE A1:NXCNC1=100PPM NOX=XXX.X

< TST TST > CAL

SETUP

SAMPLE

PRIMARY SETUP MENU

CFG

DAS RNGE PASS CLK MORE EXIT

SAMPLE

T200 NOX ANALYZER

NEXT

PREV

EXIT

Press EXIT at

any time to

return to

SETUP menu

Press EXIT at

any time to

return to the

SAMPLE display

Press NEXT of PREV to move back

and forth through the following list

of Configuration information:

MODEL NAME
SERIAL NUMBER
SOFTWARE REVISION
LIBRARY REVISION

iCHIP SOFTWARE REVISION

1

HESSEN PROTOCOL REVISION

1

ACTIVE SPECIAL SOFTWARE

OPTIONS

1

CPU TYPE
DATE FACTORY CONFIGURATION

SAVED

1

Only appears if relevant option of Feature is active.

4.6. SETUP ACAL: AUTOMATIC CALIBRATION

Instruments with one of the internal valve options installed can be set to automatically
run calibration procedures and calibration checks. These automatic procedures are
programmed using the submenus and functions found under the ACAL menu.

A menu tree showing the ACAL menu’s entire structure can be found in Appendix A-1
of this manual.

Instructions for using the ACAL feature are located in the Section 7.7 of this manual
along with all other information related to calibrating the 9110TH/M analyzer.