beautypg.com

Measurement interferences – Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 281

background image

Troubleshooting & Repair

Model 9110TH NOx Analyzer


Teledyne Analytical Instruments

261

8.2.4. MEASUREMENT INTERFERENCES

It should be d that the chemiluminescence method is subject to interferences from a
number of sources. The 9110TH/M has been successfully tested for its ability to reject
interference from most of these sources. Table 8-1 lists the most important gases, which
may interfere with the detection of NO in the 9110TH/M.

8.2.4.1. Direct Interference

Some gases can directly alter the amount of light detected by the PMT due to
chemiluminescence in the reaction cell. This can either be a gas that undergoes
chemiluminescence by reacting with O

3

in the reaction cell or a gas that reacts with

other compounds and produces excess NO upstream of the reaction cell.

8.2.4.2. Third Body Quenching

As shown in Equation 9-3, other molecules in the reaction cell can collide with the
excited NO

2

*

, preventing the chemiluminescence of Equation 9-2, a process known as

quenching. CO

2

and H

2

O are the most common quenching interferences, but N

2

and O

2

also contribute to this interference type.

Quenching is an unwanted phenomenon and the extent to which it occurs depends on the
properties of the collision partner. larger, more polarized molecules such as H

2

O and

CO

2

quench NO chemiluminescence more effectively than smaller, less polar and

electronically “harder” molecules such as N

2

and O

2

.

The influence of water vapor on the 9110TH/M measurement can be eliminated with an
optional, internal sample gas dryer. The concentrations of N

2

and O

2

are virtually

constant in ambient air measurements, hence provide a constant amount of quenching
and the interference of varying CO

2

amounts is negligible at low concentrations.

The 9110TH and 9110TM analyzers are typically used in high CO

2

concentration

environments. The pneumatic setup of these two analyzer models minimizes the
interference from CO

2

such that the analyzers conform to the standards set forth by the

US-EPA in Method 20 - NOx from Stationary Gas Turbines, available at
http://www.epa.gov/ttn/emc/promgate.html