Troubleshooting & Repair

Model 9110TH NOx Analyzer


Teledyne Analytical Instruments

289

Table 8-5: Typical Thermocouple Settings

TC

TYPE

TERMINATION

TYPE

OUTPUT

SCALE TYPE

JUMPER

BETWEEN

PINS

USED ON

JUMPER

COLOR

INPUT TC1 (J15)

K

GROUNDED

5mV / °C

2 – 12
4 – 14

9110TH/M with Mini HiCon Converter

BROWN

K

ISOLATED

5mV / °C

2 – 12
4 – 14
5 – 15

9110TH/M with Mini HiCon Converter

GREY

K

ISOLATED

10mV / °C

4 – 14
5 – 15

9110TH/M models with Moly Converter

PURPLE

J

ISOLATED

10mV / °C

1 – 11
3 – 13
5 – 15

9110TH/M models with Moly Converter

RED

J

GROUNDED

10mV / °C

1 – 11
3 – 13

9110TH/M models with Moly Converter

GREEN

8.5.7.4. Valve Control

The relay board also hosts two valve driver chips, each of which can drive up four
valves. The main valve assembly in the 9110TH/M is the NO/NO

X

- Auto-zero solenoid

valve assembly mounted right in front of the NO

2

converter housing. These two valves

are actuated with 12 V supplied from the relay board and driven by the CPU through the
I

2

C bus.

A second set of valves may be installed if the zero/span valve is enabled in the analyzer.
Specialty manifold valves may be present in the analyzer.