Model 9110TH NOx Analyzer

Troubleshooting & Repair

Teledyne Analytical Instruments

236

7.5.10.3. Status Outputs

The procedure below can be used to test the Status outputs.

V

+DC Gnd

Figure 7-15:

Typical Set Up of Status Output Test

1. Connect a cable between the “D“ pin and the “” pin on the status output

connector.

2. Connect a 1000 Ω resistor between the “+” pin and the pin for the status output that

is being tested.

3. Connect a voltmeter between the “D“ pin and the pin of the output being tested

(Table 7-10).

4. Under

the

DIAG / SIGNAL I/O menu (Section 4.13.2), scroll through the inputs and

outputs until you get to the output in question. Alternately turn the output on and off.

 The Voltmeter will read approximately 5 VDC when the output is OFF.

 The Voltmeter will read approximately 0 VDC when the output is ON.

Table 7-10: Status Outputs Pin Assignments

PIN #

STATUS

1 SYSTEM

OK

2 CONC

VALID

3 HIGH

RANGE

4 ZERO

CAL

5 SPAN

CAL

6 DIAG

MODE

7 LOW

8 SPARE