Calibration Procedures

Model 9110TH NOx Analyzer

Teledyne Analytical Instruments

171

STEP THREE

Activate NO

2

measurement stability function.

SETUP X.X

STABIL_GAS:NO2

NO NO2 NOX O2

ENTR EXIT

Press ENTR first,

then press EXIT 3

times to return to

SAMPLE menu

SAMPLE RANGE = 50.000 PPM CO =X.XXX

< TST TST > CAL

SETUP

SETUP X.X

PRIMARY SETUP MENU

CFG DAS RNGE PASS CLK

MORE

EXIT

SETUP X.X

SECONDARY SETUP MENU

COMM

VARS

DIAG ALRM

EXIT

SETUP X.X

ENTER PASSWORD:818

8

1

8

ENTR

EXIT

SETUP X.X 0) DAS_HOLD_OFF=15.0 Minutes

NEXT>

JUMP

EDIT PRNT EXIT

SETUP X.X

2) STABIL_GAS=NOX

JUMP

EDIT

PRNT EXIT

SETUP X.X

STABIL_GAS:NOX

NO

NO2

NOX

O2

ENTR EXIT

Continue pressing NEXT until ...