5.3. MANUAL CALIBRATION

The following section describes the basic method for manually calibrating the Model
9110TH/M NO

X

analyzer.

If both available DAS parameters for a specific gas type are being reported via the
instruments analog outputs e.g. NXCNC1 and NXCNC2, separate calibrations should
be carried out for each parameter.

 Use the LOW button when calibrating for NXCNC1

 Use the HIGH button when calibrating for NXCNC2.

See Section 4.13.4 for more information on analog output reporting ranges

STEP ONE:

Connect the sources of zero air and span gas as shown below.

VENT

here if input

is pressurized

Source of

SAMPLE Gas

PUMP

Instrument

Chassis

Sample

Exhaust

Span Point

Zero Air

Calibr

ated

NO

at

HIG

H

Span

Co

nc

e

nt

rat

ion

Filter

External Zero

Air Scrubbe

r

VENT if not

vented at
calibrator

MODEL T700

Gas Dilution

Calibrator

MODEL 701

Zero Gas

Generator

Figure 5-2:

Pneumatic Connections–With Zero/Span Valve Option (50A)