Analyzer id – Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual
Page 111

Operating Instructions
Model 9110TH NOx Analyzer
Teledyne Analytical Instruments
91
4.11.4. ANALYZER ID
Each type of TAI analyzer is configured with a default ID code. The default ID code for
all 9110TH/M analyzers is either 0 or 200. The ID number is only important if more
than one analyzer is connected to the same communications channel such as when
several analyzers are on the same Ethernet LAN (see Section 4.11.7); in a RS-232
multidrop chain (see Section 4.11.9) or operating over a RS-485 network (see Section
4.11.6). If two analyzers of the same model type are used on one channel, the ID codes
of one or both of the instruments needs to be changed so that they are unique to the
instruments. To edit the instrument’s ID code, press:
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
SETUP X.X
COMMUNICATIONS MENU
ID
INET
COM1
COM2
EXIT
SETUP X.
MACHINE ID: 200 ID
0 2 0 0
ENTR EXIT
Toggle these buttons
to cycle through the
available character set:
0-9
ENTR button accepts the
new settings
EXIT key ignores the new
settings
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST TST > CAL
SETUP
The ID can be any 4 digit number and can also be used to identify analyzers in any
number of ways (e.g. location numbers, company asset number, etc.)