Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual
Page 205

Calibration Procedures
Model 9110TH NOx Analyzer
Teledyne Analytical Instruments
185
To program the sample sequence shown above, follow this flow chart:
SETUP C.4 SEQ 2) ZERO–SPAN, 2:00:30
PREV NEXT MODE SET
EXIT
SAMPLE RANGE = 500.0 PPB NOX=X.X
< TST TST > CAL CALZ CZLS
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG ACAL DAS RNGE PASS CLK MORE EXIT
SETUP X.X SEQ 1) DISABLED
NEXT MODE
EXIT
SETUP X.X SEQ 2) DISABLED
PREV NEXT MODE
EXIT
SETUP X.X MODE: DISABLED
NEXT
ENTR EXIT
SETUP X.X MODE: ZERO–HI
PREV NEXT
ENTR EXIT
SETUP X.X SEQ 2) ZERO–HI, 1:00:00
PREV NEXT MODE SET
EXIT
SETUP X.X STARTING DATE: 01–JAN–02
EXIT
SETUP X.X STARTING DATE: 01–JAN–02
0 4 SEP 0 3
ENTR EXIT
Toggle to
set day,
month &
year: DD-
MON-Y
Y
SETUP X.X STARTING DATE: 04–SEP–03
EXIT
Default
value
is ON
SETUP C.4 STARTING DATE: 04–SEP–03
EXIT
Toggle to set
time: HH:MM.
This is a 24 hr
clock. PM
hours are 13-
Toggle
numbers to
set
number of
days
between
procedures
(1 367)
Toggle keys
to set
delay time for
each iteration
of the
sequence:
HH:MM
(0 – 24:00)
SETUP C.4 DELTA DAYS: 1
EXIT
SETUP C.4 DELTA DAYS: 1
0 0 2
ENTR EXIT
SETUP C.4 DELTA DAYS:2
EXIT
SETUP C.4 DELTA TIME00:00
EXIT
SETUP C.4 DELTA TIME: 00:00
0 0 :3 0
ENTR EXIT
SETUP C.4 DELTA TIME:00:30
EXIT
SETUP C.4 STARTING TIME:14:15
EXIT
EXIT returns
to the SETUP
Menu
SETUP C.4 DURATION:15.0 MINUTES
EXIT
SETUP C.4 DURATION 15.0MINUTES
3 0 .0
ENTR EXIT
SETUP C.4 DURATION:30.0 MINUTES
EXIT
Toggle keys
to set
duration for
each
iteration of
the
sequence:
Set in
Decimal
minutes
from
0.1 – 60.0
SETUP C.4 CALIBRATE: OFF
EXIT
SETUP C.4 CALIBRATE: OFF
ON
ENTR EXIT
SETUP C.4 CALIBRATE: ON
EXIT
Toggle key
between
Off and
ON
SETUP X.X TIMER ENABLE: ON
SET> EDIT
EXIT
SETUP C.4 STARTING TIME:00:00
EXIT
Toggle NEXT button until ...