Operating Instructions

Model 9110TH NOx Analyzer

Teledyne Analytical Instruments

157

LOGON 940331

940331 is the default password. To change the default password, use the variable
RS232_PASS issued as follows:

V RS232_PASS=NNNNNN

Where N is any numeral between 0 and 9.

4.15.2.8. APICOM Remote Control Program

APICOM is an easy-to-use, yet powerful interface program that allows to access and
control any of TAI main line of ambient and stack-gas instruments from a remote
connection through direct cable, modem or Ethernet. Running APICOM, a user can:

 Establish a link from a remote location to the 9110TH/M through direct

cable connection via RS-232 modem or Ethernet.

 View the instrument’s front panel and remotely access all functions that

could be accessed when standing in front of the instrument.

 Remotely edit system parameters and set points.

 Download, view, graph and save data for predictive diagnostics or data

analysis.

 Retrieve, view, edit, save and upload DAS configurations.

 Check on system parameters for trouble-shooting and quality control.

APICOM is very helpful for initial setup, data analysis, maintenance and trouble-
shooting. Figure 6-20 shows examples of APICOM’s main interface, which emulates
the look and functionality of the instruments actual front panel