beautypg.com

Warning, Wiring input – GF Signet 4632 Chlorine Dioxide Analyzer System User Manual

Page 8

background image

8

463X Chlorine System Manual

Input for
100-240 VAC

100 to 240 VAC input
Standard AC con¿ guration

12 to 24 VDC Input Wiring Conversion
• If the power source supplied to the system will be 12 to

24 VDC instead of 100 to 240 VAC, disconnect the red
and black output wires from the power supply (Figure 3)
and connect your UL approved limited-energy DC power
supply (Figure 4), using an insulated nylon parallel splice
connector such as T&B part number 2C-12 or equivalent.

• Insert input power wiring into the pre-drilled access hole

on the left side of the electrical box using the appropriate
conduit adapters to maintain the Type 4X rating.
(Customer may have to remove the supplied conduit
connectors.)

• A switch or circuit breaker rated at 5 amps DC shall be

included in the building installation.

• Install the circuit breaker in close proximity to the

equipment and within easy reach of the operator.

• Mark the circuit breaker as the disconnecting device for

the equipment.

Power Supply Terminal Identi¿ cation

Terminal

Name

Description

V. ADJ

Voltage
adjustment

Adjusts within ±10%; turning clockwise
increases output voltage

DC ON

Operation
indicator

Green LED is lit when output voltage
is on

+V, –V

DC output
terminals

+V: Positive output terminal
–V: Negative output terminal

L, N

Input
terminals

Accepts a wide range of voltages and
frequencies (100 to 240 VAC DC input)

100 to 240 VAC Input Wiring
• Open the electrical box by loosening the four corner

screws of the front cover.

• Insert input power wiring into the pre-drilled access hole

on the left side of the electrical box using appropriate
conduit adapters to maintain the Type 4X rating. (Supplied
conduit connectors may have to be removed.)

• Install the input power wires into the proper terminals

on the power supply (Figure 2). Use only 12 to 26 AWG
copper wiring.

• Recommended torque for the terminals is 7 lb-in.

NOTE: When using alternate power supply

7300-7524 (159 000 687):

• Use 10-24 AWG copper wiring, 105 ÛC, torque 4.4 lb-in.

Figure 2

System Input Power
• The panel system is pre-wired with an auto switching power

supply that is rated for 100 to 240 VAC 50/60 Hz input.

• Wire with NEC Class I, 300 volt, 105 C wire.
• A switch or circuit breaker rated at 15 amps AC shall be

included in the building installation.

• Install the circuit breaker in close proximity to the

equipment and within easy reach of the operator.

• Mark the circuit breaker as the disconnecting device for

the equipment.

• Grounding: The protective ground terminal must be

bonded to protective earth in the host equipment.

WARNING

This panel system is
wired for AC voltages
that can injure or kill.

Wiring should be performed by
quali¿ ed personnel only
.
Disconnect AC power before opening wiring enclosure.

Figure 3

+

Figure 4

Part # 7300-0024 (159 001 693) shown.
Actual power supply may differ from Figure 2.

Follow all local and government recommendations and
methods for installation of electrical connections to and
between the system and other devices.

Wiring Input

Wiring Input

This manual is related to the following products: