beautypg.com

GF Signet 4632 Chlorine Dioxide Analyzer System User Manual

Page 25

background image

25

463X Chlorine System Manual

Cl In Process
Calibration:

>

In Proc: ppm&nA
2.67 / 103.8

E X A M P L E

Calibrate Menu

DPD Cl ppm value
entered here.

ENTER

ENTER

Step 5

7

The signal level during an In-Process
Calibration must be at least 1 nA higher
than the previous Zero Calibration point.

In-Process Calibration

E X A M P L E

1. Take a water sample from the Sampling Port (after

purging it) from a stabilized and running system.

2. Use this sample to measure the chlorine content with

a colorimetric DPD test kit (not included). Refer to the
DPD kit instructions on how to perform this test.

3. Record the test results.

Example: Set the Chlorine In-Process to 2.67 ppm.

4. Go to the Calibrate Menu.
5. Scroll down ź 2 menus to the Cl In-Process menu.
6. Press Ź to enter Edit Mode.
7. Enter the chlorine reading determined from the DPD test

into the edit screen: 2.67 ppm. The Cl ppm is editable
and must be at least 0.2 ppm.

8. Press the Enter button to save the setting.
9. Exit to the View Menu.

Calibration is complete for the chlorine sensor.

NOTE: For greater accuracy, it is recommended
that the DPD test be repeated three times and
the results averaged together.

Con¿ gure the current loop and relay functions if applicable.
NOTE: The current and relay outputs can be tested in the
Options Menu.

Current Loop Settings
Current outputs are passive outputs that can be spanned in
the forward and reverse direction. Example: 0 to 5 or 5 to 0.

Example: Set a current loop source as chlorine and the
operational range to 0 to 5 ppm.

1. Go to the Calibrate Menu.
2. Scroll down ź to the Loop 1 Source menu.
3. Press Ź to enter Edit Mode.
4. Choose either the chlorine or pH sensor as the source

that will control this loop: Chlorine.

5. Press the Enter button to save the setting.

6. Scroll down ź 1 menu to the Loop 1 Rng menu.
7. Press Ź to enter Edit Mode.
8. Select the minimum and maximum process values for

the current loop output: 0 to 5 ppm.

9. Press the Enter button to save the setting.
10. Exit to the View Menu.

E X A M P L E

A diethyl-p-phenylenediamine (DPD) colorimeter test kit (not
included) is required for sensor calibration. A sample is taken
and analyzed with the DPD test kit, then this value is entered
into the Signet 8630 transmitter.

Chlorine Sensor Calibration

Calibrate Menu

Set and save

Set and save

Loop 1 Source:

Chlorine

>

Step 2

Loop 1 Rng: ppm
0.00 5.00

>

6

E X A M P L E

NOTE:
When integrated into the 463X Chlorine Panel Assembly,
the loop outputs are wired into Active outputs via the
terminal strip found inside the 463X enclosure.

Output Settings - Loops and Relays

This manual is related to the following products: