GF Signet 4632 Chlorine Dioxide Analyzer System User Manual

Page 35

background image

35

463X Chlorine System Manual

WARNING! Do not over tighten À ow cell bolts. Maximum torque is 8.1 Nm (72 Lb-In).
Over tightening the bolts can damage the À ow cell.
Do not over-tighten the bolts in an attempt to stop a leak.

Tighten the nut just enough to hold
the block together and keep the
O-rings in place.

6

Tighten the bolts in three passes to
speci¿ cation.

8

Install the nut and washers.

5

Install the remaining bolts, washers
and nuts.

7

Install the assembled pressure
regulator and the two round
O-rings into the À ow cell.

1

Install the four O-ring segments.

2

Insert the center bolt into the block.

3

4

Align and place the back block onto
the À ow cell. Check to be sure that
the O-rings are seated correctly.

9

Install the À ow cell on the panel.
Screw on the thumb nuts ¿ nger tight.
DO NOT use tools on the thumb nuts.

NOTE: It is recommended that an anti-seizing compound appropriate for
the application be used during the reassembling of the fl ow cell.

Flow Cell Assembly

1

2

3

6

5

Torque
8.1 Nm

(72 Lb-In)

MAX!

4

Tighten bolts in 3 passes

Appendix