Appendix – GF Signet 4632 Chlorine Dioxide Analyzer System User Manual
Page 33
![background image](/manuals/331162/33/background.png)
33
463X Chlorine System Manual
Sensor Removal
Figure 2
Figure 1
Under certain conditions, a dirty À ow cell and ¿ lter can create a chlorine demand
which could lower the chlorine concentration in the water À owing past the
chlorine sensor. It is recommended to clean the ¿ lter and À ow cell on a regular
basis. The frequency of necessary cleaning will depend on the application in
which the system is being used and the level of accuracy required.
Step 1. Keep the system powered on.
Step 2. Disable the relays and any output loops.
Step 3. Turn off the water À ow.
Step 4. Remove the sensors from the À ow cell. NOTE: It is not
necessary to remove the electronics from the sensor.
Step 5. Install the vinyl caps on to the sensor tips or place the tips of
the removed sensors in a cup of sample water for temporary
storage while cleaning the À ow cell.
Keep the sensor tips wet and the sensors energized.
Step 6. Remove the À ow cell from the panel by removing the
knurled nuts (Figure 1).
Step 7. Remove the remaining hardware from the À ow cell.
(Figure 2).
NOTE: It is recommended that an anti-seizing compound appropriate for the
application be used during the reassembling of the fl ow cell.
Maintenance - 463X Chlorine Analyzer Flow Cell
CAUTION: Over time, a sensor can get tight in the À ow cell ¿ tting. When removing the sensor, avoid
hitting the sensor electronics on the bottom of the wiring enclosure if the sensor suddenly releases.
Take care not to damage the components.
While holding the À ow cell, grasp the yellow electronics as a whole unit and
carefully pull upwards, gently rocking back and forth if necessary. Once the
sensor is loose, disengage the electronics, then remove the sensor.
TIP: A tool can be used to carefully pry the sensor electronics up and out if it is
dif¿ cult to remove.
Cleaning
Service the À ow cell on a clean, dirt free surface
to avoid scratches or damage to the À ow cell.
• Wash the surfaces with plain or soapy water only.
• Use gentle, liquid dishwashing soap if necessary.
• DO NOT USE commercial glass cleaning À uids.
• Use a soft, lint-free cloth.
Appendix
If your inlet pressure is less than 1 bar (15 psi), remove
the À ow regulator and quick disconnect inlet connector.
Remove ¿ lter
screen to clean.
If the ¿ lter is not
desired, it may
be removed from
the system.
Flow Regulator
Servicing the Flow Regulator and Filter
WARNING: Do not disassemble the À ow regulator.
There are no user serviceable parts inside.
Appendix