GF Signet 4632 Chlorine Dioxide Analyzer System User Manual

Page 21

background image

21

463X Chlorine System Manual

Zero Cal: ppm&nA
xxx.xx xxxx.x

View user entered Zero Calibration data in ppm and nA.

Temp at Cal:
xxxx.x °C

Temperature recorded during user In-Process Calibration.

pH at Cal:
xxx.xx pH

pH value recorded during user In-Process Calibration.

In Proc: ppm&nA
xxx.xx xxxx.x

View user In-Process Calibration value when it was entered in the Calibrate Menu.

Temp Offset:
xxxx.x °C

Temperature offset calculated from user entered temperature calibration from Calibrate Menu.

Elapsed Time:
xxxxx. hrs

Total hours of operation.

Low & High: °C
-xxxx.x +xxxx.x

Lowest and highest temperatures the Cl sensor has been subjected to during operation.

Test Relay 1:

>

Press UP or DOWN keys to manually toggle Relay 1 Off and On. The left LED on the front of
the transmitter con¿ rms operation.

Test Relay 2:

>

Press UP or DOWN keys to manually toggle Relay 2 Off and On. The right LED on the front of the
transmitter con¿ rms operation.

Read Sens Data:
No

>

If "YES" is selected the following (Read Only) screens will be shown.
If "NO" then this menu ends the Options Menu.

Cl Sensor S/N:
xxxxxxxxx

View the sensor serial number.

Cl Type & Range:
2630 xxx.x ppm

Identify the chlorine sensor type connected and its ppm range.

If Yes was chosen:

End of Options Menu