GF Signet 4632 Chlorine Dioxide Analyzer System User Manual
Page 21

21
463X Chlorine System Manual
Zero Cal: ppm&nA
xxx.xx xxxx.x
View user entered Zero Calibration data in ppm and nA.
Temp at Cal:
xxxx.x °C
Temperature recorded during user In-Process Calibration.
pH at Cal:
xxx.xx pH
pH value recorded during user In-Process Calibration.
In Proc: ppm&nA
xxx.xx xxxx.x
View user In-Process Calibration value when it was entered in the Calibrate Menu.
Temp Offset:
xxxx.x °C
Temperature offset calculated from user entered temperature calibration from Calibrate Menu.
Elapsed Time:
xxxxx. hrs
Total hours of operation.
Low & High: °C
-xxxx.x +xxxx.x
Lowest and highest temperatures the Cl sensor has been subjected to during operation.
Test Relay 1:
>
Press UP or DOWN keys to manually toggle Relay 1 Off and On. The left LED on the front of
the transmitter con¿ rms operation.
Test Relay 2:
>
Press UP or DOWN keys to manually toggle Relay 2 Off and On. The right LED on the front of the
transmitter con¿ rms operation.
Read Sens Data:
No
>
If "YES" is selected the following (Read Only) screens will be shown.
If "NO" then this menu ends the Options Menu.
Cl Sensor S/N:
xxxxxxxxx
View the sensor serial number.
Cl Type & Range:
2630 xxx.x ppm
Identify the chlorine sensor type connected and its ppm range.
If Yes was chosen:
End of Options Menu