GF Signet 4632 Chlorine Dioxide Analyzer System User Manual
Page 16
![background image](/manuals/331162/16/background.png)
16
463X Chlorine System Manual
Step 6
.
Last Cal:
Saving
Last Cal:
10-18-09
8630 Editing Procedure
Step 4
.
Last Cal:
09-18-09
>
Step 1
.
2s
(Hold)
5s
OPTIONS
menu
OR
ENTER
ENTER
Example
Step 3
.
Cl Zero
Calibration:
>
Last Cal:
09-18-09
>
Step 5
.
Last Cal:
0
9-18-09
Last Cal:
1
9
-18-09
Last Cal:
1
9
-18-09
Last Cal:
1
0
-18-09
ENTER
ENTER
CALIBRATE menu
Step 2
.
CALIBRATE:----
Enter Key Code
CALIBRATE:---
Enter Key Code
CALIBRATE:--
Enter Key Code
CALIBRATE:-
Enter Key Code
Cl Zero
Calibration:
>
Example: Calibration
Access the CALIBRATE Menu:
Press and hold the ENTER key for 2 seconds to access
the CALIBRATE menu.
Edit the À ashing element:
This is the edit mode.
The Ÿ or ź keys change the À ashing element.
The Ź key advances the À ashing element in a
continuous loop. In this example, the "Last Cal" date
was changed from 09-18-09 to 10-18-09.
All output functions remain active during editing.
Only the À ashing element can be edited.
Press ENTER to save the new value.
When you have set your desired value, pressing the
ENTER key stores the value on the screen, making it
immediately available to output functions and exits you
back to Step 3.
Enter the Key Code:
The CALIBRATE and OPTIONS menus require a
password (KEY CODE). Pressing the Ÿ, Ÿ, Ÿ, ź keys
in sequence unlocks the display and the ¿ rst menu item
will appear. If no key is pressed for 5 minutes while the
display is showing "Enter Key Code", it will return to the
VIEW menu.
Scroll the Menu:
Press the ź or Ÿ keys to scroll through the Menu.
While in this mode, pressing the Ÿ and ź keys
simultaneously will return the display to the VIEW menu.
If no key is pressed for 10 minutes, the display will
return to the VIEW menu.
Select the item to be edited:
In this example, "Last Cal" (last calibration date) is
chosen to edit. Pressing the Ź key selects the menu
item and enters the screen into edit mode.
ENTER
ENTER
ENTER
ENTER