463X Chlorine System Manual

If Pulse Mode was chosen:

Relay 2 Mode:
Off

>

Choose Relay 2 mode of operation: Off, Low, High, Window, or Pulse.
To disable this relay choose Off.

Relay 2 Source:
pH

>

Choose pH or chlorine for Relay 2. This menu screen and all subsequent Relay 2 screens
below do not appear if Relay 2 Mode is set to Off.

Relay 2 Setpnt:
0.00 pH

>

In Low or High Mode, Relay 2 will be activated when the process reaches this value.
Units of measure reÀ ect Relay 2 Source.

Relay 2 Hys:
0.20 pH

>

Relay 2 will be deactivated at Relay 2 Setpoint ± this hysteresis setting depending on High or
Low Setpoint selection.

Relay 2 Delay:
0.0 secs

>

Set the time delay for Relay 2 to activate after reaching the Setpoint.
Range: 0 to 6400 seconds.

If Low or High Mode was chosen:

Relay 1 Hys:
0.20 ppm

>

Relay 1 will be deactivated at Range setpoints ± this hysteresis setting.

Relay 1 Delay:
0.0 secs

>

Set time delay for Relay 1 to activate after reaching the setpoints set in Relay 1 Range.

Relay1 Rng: ppm
0.00 5.00

>

Enter the range where Relay 1 will activate above and below this setpoint.

Relay1 PlsRate:
120 pulses/min

>

Set the maximum pulse rate. Range: 1 to 400 pulses/min.

Relay 1 Mode:
Off

>

Choose mode of operation: Off, Low, High, Window, or Pulse. If Off, all subsequent Relay 1
functions are inactive and not visible.

Relay 1 Source:
Chlorine

>

Choose chlorine or pH for Relay 1.

Relay 1 Source:
Chlorine

>

Choose chlorine or pH for Relay 1.

Relay 1 Source:
Chlorine

>

Choose chlorine or pH for Relay 1.

Relay 1 Setpnt:
0.00 ppm

>

In Low or Hi Mode, Relay 1 will be activated when the process reaches this value.
Units of measure reÀ ect Relay 1 Source.

Relay 1 Hys:
0.20 ppm

>

Relay 1 will be deactivated at Relay 1 Setpoint ± this hysteresis setting depending on High or
Low Setpoint selection.

Relay 1 Delay:
0.0 secs

>

Set the time delay for Relay 1 to activate after reaching the Setpoint.
Range: 0 to 6400 seconds.

Relay Functions

If Low or High Mode was chosen:

If Window Mode was chosen:

Relay1 Rng: ppm
0.00 5.00

>

Enter the range where Relay 1 will activate above and below this setpoint.

Verify all relay settings if the Relay Source is changed.