beautypg.com

GF Signet 4632 Chlorine Dioxide Analyzer System User Manual

Page 28

background image

28

463X Chlorine System Manual

CAUTION!

• Remove power before wiring.
• Follow instructions carefully to avoid personal injury or

damage to the electronics.

• Refer to the wiring diagram above to connect the 2650 cable to the terminals on the 8630 Chlorine Transmitter.

• For calibration and con¿ guration please refer to the 8630 Calibrate Menu discussions above.

The electronics are pre-wired from the factory to the transmitter. Refer to the following
schematics when replacing the electronics.

8630

Chlorine

Gnd

I/O

V

+

18
17
16

Signal Ground

Digital (S

3

L) data

+5 VDC

White

Red

Black

Shield

pH

Signet 3-2750-7

(159 001 671)

N/C

8630

Chlorine

Gnd

I/O

V

+

15
14
13

Signal Ground

Digital (S

3

L) data

+5 VDC

White

Red

Black

Shield

Chlorine

Signet 3-2650-7

(159 001 670)

N/C

• The Signet 2650-7 Amperometric Electronics provide the polarization voltage and signal conditioning required by the Signet

2630-X and 2632-X Sensors.

• The Signet 2750-7 pH Electronics conditions and ampli¿ es the output from the Signet 2724-00 pH Electrode.
• Both units output a Digital (S

3

L) signal to the Signet 8630 Chlorine Transmitter.

2650-7 and 2750-7 Electronics

2650-7 Amperometric and 2750-7 pH DryLoc

®

Electronics

Wiring to the Signet 8630 Chlorine Transmitter

This manual is related to the following products: