Troubleshooting – Ohaus I150-S Indicator User Manual

Page 70

background image

70

SYMPTOM

Unit will not turn on.

Cannot rezero indicator,
or will not re-zero when
turned on.

Center of Zero display in-
dicator erratic or does not
appear with no load on
load receiver.

Cannot display weight in
desired weighing unit.

Cannot display setup pa-
rameters or weight read-
ings for second load re-
ceiver.

Unit will not change weight
reading.

REMEDY

Check power cord con-
nections. Power indica-
tor light should be lit.

Remove load on load re-
ceiver to less than en-
tered zero %.

Change allowable zero %
in “ZERO” parameter of
setup submenu.

Remove disturbances or
reduct motion.

Increase averaging level
in “A.L.” parameter of
scale submenu.

Enable desired unit (see
section on

SCALE SUB-

MENU).

Only KG and LB are avail-
able when LFT is set to
ON.

Select “duAL” for “PLAtF”
parameter of setup sub-
menu.

Check connections for
second load receiver.

Select proper load re-
ceiver.

PROBABLE CAUSE(S)

Power cord not plugged
in or properly connected.

Load on load receiver ex-
ceeds allowable zero %
entered in “ZERO” pa-
rameter of setup sub-
menu.

Retain set to ON and re-
called zero value is not
equal to current load.

Load receiver motion or
disturbances exceed cen-
ter of zero criteria.

Desired unit not set to ON
in “SELECt” parameter of
scale submenu.

“PLAtF” parameter of
setup submenu is set to
“SingLE”, or second plat-
form is not properly con-
nected.

I150 is set for dual load
receiver operation. Cor-
rect load receiver not se-
lected.

TROUBLESHOOTING

TROUBLESHOOTING

TROUBLESHOOTING

TROUBLESHOOTING

TROUBLESHOOTING