RS-232 INTERFACE

RS-232 INTERFACE

RS-232 INTERFACE

RS-232 INTERFACE

RS-232 INTERFACE

Terminal Block Connections

Terminal Block Connections

Terminal Block Connections

Terminal Block Connections

Terminal Block Connections

TARE*

external signal

PRINT*

external signal

CTS

clear to send

RD

data in

DTR

data terminal ready

TXD

data out

GND

signal ground

E GND

earth ground

* External switches may be installed as
shown. Momentary contact push button
or similar switches must be used.

RS-232 Commands

RS-232 Commands

RS-232 Commands

RS-232 Commands

RS-232 Commands

All communication is accomplished using standard ASCII format. Only the characters
shown in the following table are acknowledged by the I150. Any other commands,
control characters or spaces are ignored.

Commands sent to the indicator must be terminated with a carriage return (CR) or
carriage return, line feed (CRLF). For example, a tare command should appear as
shown in the following diagram:

TARE COMMAND - ‘T’

Field:

T

CR

LF

Length:

1

1

1