Legal for trade sense, Legal for trade, Sense – Ohaus I150-S Indicator User Manual
Page 19

19
Legal For Trade
Legal For Trade
Legal For Trade
Legal For Trade
Legal For Trade
This parameter can be set to ON or OFF.
Selecting ON automatically sets certain
parameters for legal for trade require-
ments, disables others, and limits the
overall display resolution to 1:10,000.
When this parameter is set to ON, only
load receiver #1 may be used.
To set this parameter ON or OFF:
Press
UPUP
or
DOWN
to display the
desired setting, then press
ENTER
to return to the setup submenu.
RECALIBRATION IS REQUIRED
RECALIBRATION IS REQUIRED
RECALIBRATION IS REQUIRED
RECALIBRATION IS REQUIRED
RECALIBRATION IS REQUIRED if LFT
is switched from OFF to ON and greater
than 1:10,000 resolution exists.
Sense
Sense
Sense
Sense
Sense
Sense is used to compensate for signal
losses when load receivers are located
long distances from the indicator. When
this feature is used, only load receiver #1
may be connected to the indicator. If two
load receivers are used, OFF must be
selected for this parameter.
To set the Sense parameter ON or OFF:
Press
UPUP
or
DOWN
to display the
desired setting, then press
ENTER
to return to the setup submenu.
RECALIBRATION IS REQUIRED
RECALIBRATION IS REQUIRED
RECALIBRATION IS REQUIRED
RECALIBRATION IS REQUIRED
RECALIBRATION IS REQUIRED if
SENSE parameter is changed.
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER