Graduations calibration point, Graduations, Calibration point – Ohaus I150-S Indicator User Manual
Page 22

22
Graduations
Graduations
Graduations
Graduations
Graduations
This parameter specifies the increments
in which the last digit of the display will
count, and sets the decimal point posi-
tion. Graduations can be set to a decimal
multiple of 1, 2, or 5 (i.e. .02, .2, 2, 20,
etc.).
To set graduations:
Enter the desired number (with deci-
mal point) using the numeric keypad,
then press
ENTER
to return to the
setup submenu.
RECALIBRATION IS REQUIRED
RECALIBRATION IS REQUIRED
RECALIBRATION IS REQUIRED
RECALIBRATION IS REQUIRED
RECALIBRATION IS REQUIRED if the
graduations setting is changed. Setup
submenu changes are not saved if you
exit menus without recalibrating.
LFT NOTE:
LFT NOTE:
LFT NOTE:
LFT NOTE:
LFT NOTE: Resolution is limited to
1:10,000 for Legal for Trade installations.
Calibration Point
Calibration Point
Calibration Point
Calibration Point
Calibration Point
The calibration point is the weight value
that will be used to calibrate the indicator.
It can be any value from 10% to 100% of
the Full Scale Point and must be entered
in the same weighing unit selected for the
calibration unit.
To enter the calibration point:
Enter the desired value using the
numeric keypad, then press
ENTER
to return to the setup submenu.
RECALIBRATION IS REQUIRED
RECALIBRATION IS REQUIRED
RECALIBRATION IS REQUIRED
RECALIBRATION IS REQUIRED
RECALIBRATION IS REQUIRED if the
calibration point is changed. Setup sub-
menu changes are not saved if you exit
menus without recalibrating.