Graduations

Graduations

Graduations

Graduations

Graduations

This parameter specifies the increments
in which the last digit of the display will
count, and sets the decimal point posi-
tion. Graduations can be set to a decimal
multiple of 1, 2, or 5 (i.e. .02, .2, 2, 20,
etc.).

To set graduations:

Enter the desired number (with deci-

mal point) using the numeric keypad,

then press

ENTER

to return to the

setup submenu.

RECALIBRATION IS REQUIRED

RECALIBRATION IS REQUIRED

RECALIBRATION IS REQUIRED

RECALIBRATION IS REQUIRED

RECALIBRATION IS REQUIRED if the
graduations setting is changed. Setup
submenu changes are not saved if you
exit menus without recalibrating.

LFT NOTE:

LFT NOTE:

LFT NOTE:

LFT NOTE:

LFT NOTE: Resolution is limited to
1:10,000 for Legal for Trade installations.

Calibration Point

Calibration Point

Calibration Point

Calibration Point

Calibration Point

The calibration point is the weight value
that will be used to calibrate the indicator.
It can be any value from 10% to 100% of
the Full Scale Point and must be entered
in the same weighing unit selected for the
calibration unit.

To enter the calibration point:

Enter the desired value using the

numeric keypad, then press

ENTER

to return to the setup submenu.

RECALIBRATION IS REQUIRED

RECALIBRATION IS REQUIRED

RECALIBRATION IS REQUIRED

RECALIBRATION IS REQUIRED

RECALIBRATION IS REQUIRED if the
calibration point is changed. Setup sub-
menu changes are not saved if you exit
menus without recalibrating.