7.

Apply the required load to the load

receiver, and then press

ENTER

.

8.

The display will show “buSY” while
the I150 reads the last calibration
point.

9.

“CALib” will be displayed when cali-
bration is completed and the calibra-
tion submenu will be returned.

If two load receivers are being used,

you can press

DOWN

to switch to the

second load receiver, and repeat the

procedure.

NOTE: If calibration is aborted at any
time, previous calibration data will re-
main.

Calibration can be aborted by pressing

CLRR

or

OFFF

.

End

End

End

End

End

You must select “End” to exit the calibra-
tion submenu.

1.

Press

ENTR

when “End” is displayed.

2.

“CAL” will be displayed and the main
menu will be returned.

NOTE: Calibration data is permanently
stored in memory when you use “End” to
exit from the Main menu, or when the OFF
key is pressed.