Reset to Factory Defaults

Reset to Factory Defaults

Reset to Factory Defaults

Reset to Factory Defaults

Reset to Factory Defaults

This parameter is used to reset all scale
submenu parameters to the factory de-
fault settings outlined in the following table.

To reset to the factory defaults,

press

ENTER

when “rESEt” is displayed.

Averaging Level

Averaging Level

Averaging Level

Averaging Level

Averaging Level

Averaging level compensates for vibra-
tion or excessive air currents on the load
receiver. During operation the I150 con-
tinually takes weight readings from the
load cell. Successive readings are then
averaged to achieve a stabilized display.
This parameter specifies whether the I150
will average more or fewer successive
readings. Select one of four averaging
levels.

To select an averaging level:

Enter the desired level (0, 1, 2, or 3)

using the numeric keypad, then press

ENTER

to return to the scale sub-

menu.

Minimum

Averaging

Maximum

Averaging

PARAMETER

LFT = OFF

LFT = ON

(Menu changes

(Menu changes

disabled w/jumper)

enabled)

Averaging Level

3

3

3

Stability Range

0.5 grad

unchanged

0.5 grad

Unit Selection

all units = ON

lb, kg = ON

lb, kg = ON

Auto Zero Tracking

0.5 grad

unchanged

0.5 grad

Custom Unit Exponent

0

N/A

N/A

Custom Unit Mantissa

1.0

N/A

N/A

Average Display

1

N/A

N/A

PC Error Level

100%

N/A

N/A

Set Point

OFF

OFF

OFF

Store

Tare

Tare

Tare

Retain

OFF

OFF

OFF

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER