Linearization Point
Linearization Point
Linearization Point
Linearization Point
Linearization Point
This parameter is used when 3-point lin-
earization is required. The linearization
point can be set to either 1/2 of the
Calibration point, or “NONE” to disable 3-
point linearization.
To set the linearization point:
Press
UPUP
or
DOWN
to display the
desired setting, press
ENTER
to re-
turn to the setup submenu.
LFT NOTE:
LFT NOTE:
LFT NOTE:
LFT NOTE:
LFT NOTE: Parameter does not appear
on menu when LFT is set to ON.
RECALIBRATION IS REQUIRED
RECALIBRATION IS REQUIRED
RECALIBRATION IS REQUIRED
RECALIBRATION IS REQUIRED
RECALIBRATION IS REQUIRED if the
linearization point is changed. Setup sub-
menu changes are not saved if you exit
menus without recalibrating.
End
End
End
End
End
You must select “End” to exit the Setup
submenu. To exit the setup submenu:
1.
Press
ENTE
when “End” is displayed.
2.
“SEtUP” will be displayed and the
main menu will be returned.
NOTE: Parameter settings are perma-
nently stored in memory when you use
“End” to exit from the Main menu, or when
the OFF key is pressed. However, Setup
submenu parameter changes require re-
calibration in order to be saved.