Calibration unit full scale point, Calibration unit, Full scale point – Ohaus I150-S Indicator User Manual
Page 21

21
Calibration Unit
Calibration Unit
Calibration Unit
Calibration Unit
Calibration Unit
The indicator can be calibrated using
pounds or kilograms; no other units are
acceptable. If dual load receivers are
used, the same calibration unit applies to
both.
When you access this parameter, the
display will show the word “Unit” and the
current calibration unit indicator.
To change the calibration unit:
Press
UPUP
or
DOWN
to switch to
the desired unit, then press
ENTER
to return to the setup submenu
RECALIBRATION IS REQUIRED
RECALIBRATION IS REQUIRED
RECALIBRATION IS REQUIRED
RECALIBRATION IS REQUIRED
RECALIBRATION IS REQUIRED if the
calibration unit is changed. Setup sub-
menu changes are not saved if you exit
menus without recalibrating.
Full Scale Point
Full Scale Point
Full Scale Point
Full Scale Point
Full Scale Point
The maximum capacity of the load re-
ceiver is entered in this parameter using
the same weighing unit selected for the
calibration unit.
To enter the full scale point:
Enter the number using the numeric
keypad, then press
ENTER
to return
to the setup submenu.
RECALIBRATION IS REQUIRED
RECALIBRATION IS REQUIRED
RECALIBRATION IS REQUIRED
RECALIBRATION IS REQUIRED
RECALIBRATION IS REQUIRED if the
full scale point is changed. Setup sub-
menu changes are not saved if you exit
menus without recalibrating.
LFT NOTE:
LFT NOTE:
LFT NOTE:
LFT NOTE:
LFT NOTE: Resolution is limited to
1:10,000 for Legal for Trade installations.
Enter Full Scale Point
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER