Number of load reeceivers zero, Number of load receivers, Zero – Ohaus I150-S Indicator User Manual
Page 20

20
Number of Load Receivers
Number of Load Receivers
Number of Load Receivers
Number of Load Receivers
Number of Load Receivers
The I150 is configured for operation with
one (single) or two (dual) load receivers
using this parameter:
Press
UPUP
or
DOWN
to display the
desired setting, then press
ENTER
to return to the setup submenu.
If Dual is selected:
If Dual is selected:
If Dual is selected:
If Dual is selected:
If Dual is selected:
If dual is selected, the load receiver indi-
cators on the display will light to show
which load receiver is currently being
displayed. Some parameters can be set
separately for load receiver 1 and 2.
To switch from one to the other,
press
UPUP
or
DOWN
.
LFT NOTE:
LFT NOTE:
LFT NOTE:
LFT NOTE:
LFT NOTE: “SingLE” is automatically se-
lected when LFT is set to ON.
Zero
Zero
Zero
Zero
Zero
This parameter specifies the percentage
of full capacity load that may be cleared
by pressing ZERO. Zero limits may be
required in certain applications. The value
can be set to any number from 2% to
100%.
To set the allowable Zero:
Enter the desired percentage (from 2
to 100) using the numeric keypad,
then press
ENTER
to return to the
setup submenu.
LFT NOTE:
LFT NOTE:
LFT NOTE:
LFT NOTE:
LFT NOTE: Must be set to
≤
4% zero for
Canadian Legal for Trade installations.
Load Receiver #2
Load Receiver #1
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER