Performance Assurance Checks

13-20 Maintenance

M

a

int

e

na

nc

e

Performing
the Blood
Analysis
Module Self-
Test

Auto Check

- The Blood Analysis Module performs a check when the

system is switched on

Self-Test

- The Self-Test with the External Simulator M3634A may be

performed at any time.

TEST - Insert the External Simulator into the cartridge slot of the
module.

RESULT - Either PASS or FAIL will be displayed on the Monitor to
indicate whether the module has passed the self-test or not. If the
module fails the self-test, try the procedure again. If it fails again, wait
half an hour and then try the procedure again. If the module continues
to fail the self-test the module should be replaced.

Performing
the
Recorder
Module Self-
Test

Auto Check

- The recorder module performs a check when it is plugged

into the parameter module rack.

Self-Test

- The Recorder Module check may be performed at any time.

TEST - Turn on the system power and plug in the recorder module

RESULT - The LED on the front panel indicates whether the module has
passed the self-test by flashing.

LED flashes twice - module has passed check LED flashes once - module
failed self-test. Take the module out and try the procedure again to
make sure there is a good connection. LED does not flash - module
processors are not working properly and the module should be
replaced

Performing
the Data
Manage-
ment Data-
base Self-
Test

The data management database is tested during system start-up with an
SRAM checksum procedure. If there is an error found in the SRAM
checksum, the system will restart (cold start) and all patient data will be
lost.