Controls and Connectors

11-26 Monitor Installation and Patient Safety

M

onitor

Ins

ta

lla

ti

on

a

nd Pa

ti

e

n

t Sa

fe

ty

2. V AMP; this is the vertical amplitude control used to adjust the

height of the displayed video.

3. V SHIFT; this is the vertical shift control used to adjust the vertical

position of the displayed video.

4. H AMP; this is the horizontal amplitude control used to adjust the

width of the displayed video.

5. EAST WEST; this control is used to adjust the pin-cushion effect of

the displayed video.

Connectors of

the M1092A,
M1094A and

M1094B

Displays

1. There are two fuse holders for the line protection fuses. The fuses

depend on the type of monitor and the power supply as
summarized below:

2. Local Power Connector; a 3-pin connector is used to input the

local line voltage.

3. System Power Connector; a 15-pin “D” type connector is used to

output 60V DC (100W max) to a similar connector on the DC/DC
converter in the rear of the computer module.

4. Video In Connector; a 9-pin “D” type connector is used to input the

video signal to be displayed from the display controller in the
computer module (DSPC for monochrome or CDSPC-Video for
color).

5. Video Out Connector; a 9-pin “D” type connector is used to output

the video signal to be displayed on a duplicate display.

200-240V M1094A color monitor

2.5A 250V

200-240V M1094B color monitor

3.15A 250V

200-240V monochrome monitor

2A 250V

100-120V color or monochrome monitor

4A 125V
(Slow-Blow)