Using Philips Patient Care System with an Arrhythmia Computer
Other Patients 4-11
Ot
h
e
r Pa
ti
e
n
ts
R-on-T VPBs (**)
For HR < 100, a VPB with R-R interval <
1/3 second and< 1/3 of the average R-R
interval followed by a compensatory
pause, or 2 such Vs without a
compensatory pause in 5 min. (When HR
> 100, 1/3 R-R interval is too short for
detection.
VENT BIGEMINY (**)
Dominant rhythm of SV, V, SV,V.
VENT TRIGEMINY (**)
Dominant rhythm of SV, SV, V, SV, SV, V.
MULTIFORM VPBs (**)
The occurrence of two different shaped
VPBs within last 300 beats, repeated in
the last 60 beats.
HR XXX >UUU (**)
Heart-rate >UUU, where UUU is the
adjustable bedside high alarm limit. The
heart-rate of XXX is the highest heart-rate
measured since the alarm was detected.
HR XXX < LLL(**)
HR < LLL, where LLL is the adjustable
bedside low heart-rate limit. The heart-
rate of XXX is the lowest heart-rate
measured since the alarm was detected.
IRREGULAR (HR) (**)
Irregular SV rhythm