Using Philips Patient Care System with an Arrhythmia Computer

Other Patients 4-11

Ot

h

e

r Pa

ti

e

n

ts

R-on-T VPBs (**)

For HR < 100, a VPB with R-R interval <
1/3 second and< 1/3 of the average R-R
interval followed by a compensatory
pause, or 2 such Vs without a
compensatory pause in 5 min. (When HR
> 100, 1/3 R-R interval is too short for
detection.

VENT BIGEMINY (**)

Dominant rhythm of SV, V, SV,V.

VENT TRIGEMINY (**)

Dominant rhythm of SV, SV, V, SV, SV, V.

MULTIFORM VPBs (**)

The occurrence of two different shaped
VPBs within last 300 beats, repeated in
the last 60 beats.

HR XXX >UUU (**)

Heart-rate >UUU, where UUU is the
adjustable bedside high alarm limit. The
heart-rate of XXX is the highest heart-rate
measured since the alarm was detected.

HR XXX < LLL(**)

HR < LLL, where LLL is the adjustable
bedside low heart-rate limit. The heart-
rate of XXX is the lowest heart-rate
measured since the alarm was detected.

IRREGULAR (HR) (**)

Irregular SV rhythm

Alarm

Minimum Condition Required for
Alarm