beautypg.com

Making calibrated ecg recordings, If the recorder is busy – Philips V24CT User Manual

Page 195

background image

Realtime Wave Recordings

Recording Functions 6-23

Re

c

o

rd

ing

Func

tions

Note—

(CMS only):

The number of recording channels available depends

on the recorder. If you change from a 4- to a 2-channel recorder, only the
first two waves are recorded.

Note—

Non-preset recordings are made on the same recorder as procedure

recordings (selected in the Procedure Recordings Task Window). If the
power is off for more than 1 minute, a non-preset recording automatically
reverts to the default waves and recorder for Mode A. However, the speed
(25 mm/s), waves non-overlapped, and continuous runtime do not change.

Note—

If a realtime recording is requested during an alarm recording, it is

produced when the alarm recording is finished. If a realtime recording is
in progress, and another parameter goes into alarm, an alarm message
appears on the realtime recording. When the realtime recording is
finished, the alarm recording is produced.

Making
Calibrated
ECG
Recordings

If you want your ECG recording to be calibrated, you must select the size
of the ECG wave, depending on your recorder sector size. To select the
size, refer to the “Switching between EASI and Standard leads” on page
14-22.
The calibrated ECG wave recording will be 10mm/mV.

If the
Recorder is
Busy

Realtime:

If a delayed recording, or any recording from another monitor,

is in progress at the central station at the time of your request, you are
alerted with a message on the monitor display, and the request is queued.

When the recorder is free, the recording runs. It consists of current
waveforms (not the waveforms from the time the recording was
requested).

If the recorder is busy with a realtime recording and you request another
realtime recording from the same monitor, the current recording stops
and the newly requested recording starts.

Delayed:

If the recorder is busy at the time of your request, the waveform

data

and the request are queued.

When the recorder is free, the recording runs. The recording shows the
waveforms from the time the recording was requested.