Using Philips Patient Care System with an Arrhythmia Computer

4-14 Other Patients

Ot

h

e

r Pa

ti

e

n

ts

** R-ON-T PVCs

For HR <100, a PVC with R-R interval <1/3
the average interval followed by a
compensatory pause of 1.25 x average R-R
interval or 2 such Vs without a
compensatory pause occurring within 5 min.
of each other. (When HR >100, 1/3 R-R
interval is too short for detection.)

** VENT BIGEMINY

A dominant rhythm of N, V, N, V
(N=supraventricular beat, V=ventricular
beat)

** VENT TRIGEMINY

A dominant rhythm of N, N, V, N, N, V
(N=supraventricular beat, V=ventricular
beat)

** PVCs > xxx/MIN

PVCs within one minute exceeded the PVCs
/min limit (xxx)

** MULTIFORM PVCs

The occurrence of two differently shaped
Vs, each occurring at least twice within the
last 300 beats as well as each occurring at
least once within the last 60 beats

** HR yyy > xxx

Heart Rate (yyy) greater than the upper HR
limit (xxx)

** HR yyy < xxx

Heart Rate (yyy) lower than the lower HR
limit (xxx)

** IRREGULAR HR

Consistently irregular rhythm (irregular R-R
intervals)

Message

Minimum Condition Requiring an Alarm