background image

oxyCRG Display

3-26 Setting up your Monitor

S

e

tt

ing up y

our

M

onito

r

Approximately the bottom two thirds of the Main Screen area are
covered with the three oxyCRG waveforms.

The number of realtime waves that you can display simultaneously with
oxyCRG is related to the maximum number of realtime waves that your
system is capable of displaying. Details are provided in the following
table:

Main Screen

Notes on
oxyCRG

The oxyCRG information is overlayed on the realtime waves at the
bottom of the Main Screen. These realtime waves disappear from
the display, but the parameters are still being monitored, and the
parameter alarms are still active.

If oxyCRG covers part of an overlap channel, the entire overlap
channel disappears from the display.

The

softkey has no effect on the oxyCRG section

of the Main Screen, because the display speed is very low.

The beat-to-beat heart rate (btbHR) in the oxyCRG display will
vary from the heart rate (HR) numeric on the Main Screen, since
the HR numeric is an averaged value.

If the monitor is switched off for less than 60 seconds, oxyCRG
data is retained. However, when the power is switched back on,
you will see gaps in the oxyCRG display. The gaps correspond to
the amount of time it takes the monitor to get the first readings
after power is restored, not the amount of time that the monitor
was switched off. For this reason, you cannot properly measure
the time difference between an event which occurs before power

Standard Display

oxyCRG Display

4 realtime waves

oxyCRG + 1 realtime wave

6 realtime waves

oxyCRG + 2 realtime waves

8 realtime waves

oxyCRG + 3 realtime waves

Freeze Waves