Neonatal Event Review Recordings

Recording Functions 6-33

Re

c

o

rd

ing

Func

tions

Neonatal Event Review Recordings

Tabular
Neonatal
Event
Recordings

You can record the contents of the Event Review task window in a
tabular format on an M1116A/B plug-in recorder. The recorder selection
must be set to “PLUG-IN” (the default) in the Procedure Recordings task
window before neonatal events can be recorded.

You can access the Procedure Recordings task window by selecting:

on the control panel and

in the Monitor Setup selection window.

Note—

Neonatal Event Review recordings can only be made on the

M1116B plug-in recorder. The M1116A plug-in recorder, the M1117A four
channel recorder, and the central recorder do not support Neonatal
Event Review recordings.

Header Information

Each neonatal event recording begins with general information such as
the recording date, time, patient’s name, medical record number and the
bed label. Beneath this, comes the Event Summary information including
the total number of stored events and their respective event categories.
See the “Neonatal Event Review” section of this manual (Chapter 9,
Volume 1) for further details on event categories.

Neonatal Event Data

A tabular recording of neonatal event data includes the following
information:

1. The date and time the event was captured

2. The event labels (Brady, Desat and Apnea Duration). For reference

purposes, the event labels are reprinted after every 8 columns. For
filing purposes, the recording strip can be cut after every 8 columns
and attached to a standard sized sheet of paper.

3. The trigger event, which is underlined