beautypg.com

Philips V24CT User Manual

Page 122

background image

Changing the Patient Category

3-48 Setting up your Monitor

S

e

tt

ing up y

our

M

onito

r

These safety values effect the measurement range of each category. The
Adult category has the greatest range (systolic 30 to 270 mmHg) and the
Neonate has the smallest (systolic 30 to 130 mmHg). Therefore you
cannot use Neonate or Pediatric if the expected pressure values
are outside the measurement range for that category.

The following tables list the measurement ranges and safety values for
each patient category.

NBP Measurement Range (in mmHg) for Differing Patient
Categories

Measurement Range

Patient Category

Adult

Pedi

Neo

Systolic

30 to 270

30 to 180

30 to 130

Mean

20 to 255

20 to 160

20 to 120

Diastolic

10 to 245

10 to 150

10 to 100

Cuff Size

A to E

Any Cuff

1 to 4 & A

Default Inflation Pressure

165 mmHg

125 mmHg

100 mmHg

NBP Safety Values for Differing Patient Categories

Safety Value

Patient Category

Adult

Pedi

Neo

Continuous Pressure

15 mmHg
(for 120s)

5 mmHg
(for 120s)

5 mmHg (for
60s)

Time to inflate
from 0 to 60 mm Hg

40s

20s

20s

Max. inflation time

60s

40s

30s