General cleaning of the System

13-2 Maintenance

M

a

int

e

na

nc

e

General cleaning of the System

The patient monitor should be kept free of dust and dirt.

Exterior cleaning of the case and screen is recommended. Clean it with
a lint-free cloth, moistened with either warm water (40°C/104°F. max)
and soap, a diluted non-caustic detergent or one of the approved
cleaning agents listed below:

Caution

To avoid damage to the product, observe the following general
precautions for cleaning. You should only

deviate when this is explicitly

described in the cleaning instruction for the individual transducer or
accessory.

1. Do not use strong solvents such as acetone or trichloroethylene.

2. Always dilute according to the manufacturers instructions, or use

lowest possible concentration.

3. Never use abrasive material (such as steel wool or silver polish).

4. Do not allow liquid to enter into the product.

5. Never submerge any part of the system.

6. Do not pour liquid onto the system during cleaning.

Soaps

mild soaps

Tensides

(dishwasher

detergents)

Edisonite Schnellreiniger

, Alconox

Ammonias

Dilution of Ammonia <3%, Window
cleaner

Alcohol

Ethanol 70%, Isopropanol 70%, Window
cleaner