Using Philips Patient Care System with an Arrhythmia Computer

4-10 Other Patients

Ot

h

e

r Pa

ti

e

n

ts

Arrhythmia Alarms on the 78560A and 78720 Arrhythmia
Computers

Alarm

Minimum Condition Required for
Alarm

ASYSTOLE (***)

No QRS for 4 consecutive seconds in
absence of V fib or chaotic signal.

VENT FIB/TACH (***)|

Fibrillatory waveform detected for 4
consecutive seconds.

HR BRADY XXX < BBB (***) Heart rate < lower HR limit set or 40 bpm

whichever is lower.

VENT TACHY (***)

Either VENT RHYTHM with average
heart-rate > tachy HR limit or RUN VPB's
> tachy run limit with rate for the run >
tachy HR limit.
Tachy run limit is adjustable from 3 to 9
using the Change Limits display. Tachy
HR limit is adjustable from 80 bpm to 120
bpm using the Change Limits display.

VENT RHYTHM (**)

Dominant rhythm of adjacent V's and HR
< = 120

PAIR VPBs (**)

Two consecutive VPBs between non
VPBs

PACER NON-CAPT/PACER
NOT-CAPTURE (**)

Missed beat with pace pulse in paced
patient

MISSED BEAT(S) (**)

Beat omitted with HR < 120 or no beat
for 1 second with HR >120