beautypg.com

Philips V24CT User Manual

Page 338

background image

Controls and Connectors

11-22 Monitor Installation and Patient Safety

M

onitor

Ins

ta

lla

ti

on

a

nd Pa

ti

e

n

t Sa

fe

ty

Note—

Do not apply a voltage of more than

±

12V to any of these outputs.

The connectors on the rear panel of the computer module (together with
their connecting function cards) are:

1. System Power Connector; on the M1046A Computer Module, this

is a connector used to input the 60V dc power supply from the
CRT display module. It is not for use on the M1046B Computer
Module but is instead fitted with a stub connector.

2. Equipotential Grounding Terminal; this is used to connect the

computer module to the hospital´s grounding system.

3. Philips monitoring network connectors (Philips Network

Interface); these are Philips monitoring network connectors, used
to input/output information to/from a Philips monitoring network.

4. Human Interface Connector (Utility CPU); this is a Philips HIL

connector, used to input information from the keypad in the
display module.

5. Nurse Paging Relay output; this is a mini-phone jack, used to

connect to the hospital´s Nurse Paging System.

6. Analog Interface Connector; a 37-pin “D” type connector is used to

output information to analog devices.

7. RS232 Connectors (RS232 Interface); these are 25-pin “D” type

connectors, used to output information to printers or computers,

RS232 Connectors

±

12 V

HDLC Connector

5 V

Video Out (9)

5 V

Flatscreen Remote Power Supply On/Off

15 V

Video Out (11)

60 V

Maximum Voltages

This manual is related to the following products: