The v24 and v26 connectors, The v24 and v26 connectors -34 – Philips V24CT User Manual
Page 350

Controls and Connectors
11-34 Monitor Installation and Patient Safety
M
onitor
Ins
ta
lla
ti
on
a
nd Pa
ti
e
n
t Sa
fe
ty
The V24 and
V26
Connectors
The monitor has several connectors which are shown below.
Do not apply a voltage of more than +12V to any of the pins on the
defibrillator sync, parameter module rack, Philips monitoring network
(SDN) or equipotential ground connectors.
The secondary ground complies with IEC 601-1-1 as shown on page 10-4.
The connectors on the monitor are:
1. AC Power Connector; this connector is used to connect the
monitor to an AC power source.
Note—
If you have a Philips M1026A Anesthetic Gas Module (AGM), do
not daisy-chain the AC power through the AGM. This configuration may
degrade RFI performance.
2. Secondary Ground Screw; this is a posidrive screw to connect the
monitor chassis to earth ground per IEC 601-1-1 standard.
3. Equipotential Ground Post, this connects the monitor to the
equipotential ground system.
AC LINE INPUT
RS232 CONNECTOR
(Rear Housing with
Serial Prefix <3608)
PROTECTIVE
RS232 CONNECTOR
(Rear Housing with
Serial Prefix >=3608)
EQUIPOTENTIAL GROUNDING
DEFIB
CONNECTOR
RACK
CONNECTOR
marge3.tif
EARTH GROUND
PHILIPS MONITORING NETWORK (SDN)
SINGLE CONNECTOR