The V24 and
V26
Connectors

The monitor has several connectors which are shown below.

Do not apply a voltage of more than +12V to any of the pins on the
defibrillator sync, parameter module rack, Philips monitoring network
(SDN) or equipotential ground connectors.

The secondary ground complies with IEC 601-1-1 as shown on page 10-4.

The connectors on the monitor are:

1. AC Power Connector; this connector is used to connect the

monitor to an AC power source.

Note—

If you have a Philips M1026A Anesthetic Gas Module (AGM), do

not daisy-chain the AC power through the AGM. This configuration may
degrade RFI performance.

2. Secondary Ground Screw; this is a posidrive screw to connect the

monitor chassis to earth ground per IEC 601-1-1 standard.

3. Equipotential Ground Post, this connects the monitor to the

equipotential ground system.

AC LINE INPUT

RS232 CONNECTOR

(Rear Housing with

Serial Prefix <3608)

PROTECTIVE

RS232 CONNECTOR

(Rear Housing with

Serial Prefix >=3608)

EQUIPOTENTIAL GROUNDING

DEFIB

CONNECTOR

RACK

CONNECTOR

marge3.tif

EARTH GROUND

PHILIPS MONITORING NETWORK (SDN)

SINGLE CONNECTOR