Other Functions You Can Configure
Setting up your Monitor 3-33
S
e
tt
ing up y
our
M
onit
o
r
Other Functions You Can Configure
Parameters On/Off
Individual parameters can be switched ON or OFF on the monitor. No
alarms or INOPs are reported for parameters that are switched off, and
they are not available for display, trending, or recording on the monitor.
1. Press the softkey
to highlight the required
parameter. To view parameters which are not displayed, press the
softkey
.
2. Press the softkey
to set the selected parameter on
or off. Repeat this procedure for each parameter you want to set.
3. Press the softkey
if you want to preset parameter
settings when the module is plugged in WITHOUT the transducer.
Switch this setup off again using
or
.
Module Setup
Parametr On/Off
Select Parametr
Next Page
On/Off Parametr
On/Off Setup
On/Off Setup
Silence/Reset