Configuring
the Other
Patients
Controls

With the Other Patients Controls you can place your own bed into one of
eight groups: A, B, C, D, E, F, G or the ALL group, based on nursing care
assignments within a unit.

The Other Patients Controls Task Window allows you to change the Send/
Receive alarms configuration for your bed.

Note—

When a bed is assigned to the ALL group, alarms can be received

from and sent to beds in group A-G as well as the other beds in the ALL
group. When a bed is assigned to any group in A-G, alarms are received
from and sent to any bed in the same group and also to beds in the ALL
group.

Note—

The bed will automatically be assigned to the ALL group if the

connection to the Philips monitoring network is interrupted (for example,
if the Philips monitoring network cable at the back of the monitor is
disconnected, or the central station is switched off) and the monitor
comes back on a different branch.

Note—

Any changes made to Other Patients settings will be stored for up

to three hours after the monitor is switched off.