Philips V24CT User Manual
Page 134

The Test Signals Function
3-60 Setting up your Monitor
S
e
tt
ing up y
our
M
onito
r
System Self-Test Values:
Module
Test Numeric Limits
Test Waveforms
ECG and ECG/
RESP
ECG
100 bpm in ADULT mode
125 bpm in NEO/PEDI mode
Simulated square wave and numeric
RESP
15 rpm in ADULT mode
30 rpm in PEDI mode
55 rpm in NEO mode
Simulated Resp wave and numeric
Pressure
Simulated square wave and numerics
Systolic
120 mmHg ADULT
60 mmHg PEDI/NEO
Diastolic
0 mmHg ADULT/PEDI/NEO
SpO
2
/Pleth
SpO
2
100%
Numeric only
Pleth
Wave on display
Simulated Wave
tcpO
2
/tcpCO
2
Numeric only
tcpO
2
60 mmHg (8.0 kPa)
tcpCO
2
40 mmHg (5.3 kPa)
CO
2
Simulated square wave and numerics
ETCO
2
40 mmHg (6.0 kPa)
IMCO
2
0 mmHg (0.0 kPa)
AWRR
25 rpm
Temperature
40°C (104°F)
Numeric only