The Test Signals Function

3-60 Setting up your Monitor

S

e

tt

ing up y

our

M

onito

r

System Self-Test Values:

Module

Test Numeric Limits

Test Waveforms

ECG and ECG/
RESP

ECG

100 bpm in ADULT mode
125 bpm in NEO/PEDI mode

Simulated square wave and numeric

RESP

15 rpm in ADULT mode
30 rpm in PEDI mode
55 rpm in NEO mode

Simulated Resp wave and numeric

Pressure

Simulated square wave and numerics

Systolic

120 mmHg ADULT
60 mmHg PEDI/NEO

Diastolic

0 mmHg ADULT/PEDI/NEO

SpO

2

/Pleth

SpO

2

100%

Numeric only

Pleth

Wave on display

Simulated Wave

tcpO

2

/tcpCO

2

Numeric only

tcpO

2

60 mmHg (8.0 kPa)

tcpCO

2

40 mmHg (5.3 kPa)

CO

2

Simulated square wave and numerics

ETCO

2

40 mmHg (6.0 kPa)

IMCO

2

0 mmHg (0.0 kPa)

AWRR

25 rpm

Temperature

40°C (104°F)

Numeric only