Alarm Setup

Alarm Functions 5-15

Ala

rm Func

tions

The Nurse
Call Relay

Note—

The nurse call relay is not available for the V24CT and V26CT.

The nurse call relay is a paging system which warns the nurse if an alarm
occurs on a monitor. It becomes active in one of four situations
depending on the way it has been configured:

Red only alarms

Red and Yellow alarms

Red and INOP alarms

Red and Yellow INOP alarms.

The nurse call relay is not active when the monitor’s alarm capability is
switched off.

Note—

The delay time to the nurse call relay output is max. 1 sec.

When an alarm has been silenced using the

key, and the

audible 2 or 3 minute alarm reminder is configured to be either On or Re-
alarm, the nurse call relay signals will behave as follows:

Alarm

Reminder

Reminder

Time

Nurse Call Relay Signal Behavior

On

2 or 3 mins

The nurse call relay signals are repeated
for a duration of 6 seconds every 2 or 3
minutes unless the alarm condition is
eliminated

Re-alarm

2 or 3 mins

The nurse call relay signals are activated
after 2 or 3 minutes. The alarm tone and
relay signals run continuously until they
are interrupted by pressing the

key.

Silence/Reset

Silence/Reset