Performance Assurance Checks

Maintenance 13-15

M

a

int

e

na

nc

e

Performing
the ECG
Module and
ECG/RESP
Self-Test

Auto Check

- None

Self-Test

The ECG and ECG/RESP self-test may be performed at any

time as the lead selector and pace pulse detection are inhibited during the
check. The module will produce ECG test signals at the beginning of the
processing circuitry, and process the test signals in the same way patient
signals are processed. The test signals last about 30 seconds and then the
display returns to the normal monitoring mode.

TEST - Press

.

RESULT - If the module is functioning correctly, the following heart rate
and respiration are displayed without the alarms being activated:

ECG 100 bpm in ADULT mode 125 bpm in NEO/PEDI mode

RESP 15 rpm in ADULT mode 30 rpm in PEDI mode 55 rpm in NEO mode

Cardiac Output

5 l/min (+0.1 l/min)
(Blood Temp: 37°C)
(Inj Temp: 0°C)

No wave

tcpO

2

/tcpCO

2

No wave

tcpO

2

60 mmHg (8.0 kPa)

tcpCO

2

40 mmHg (5.3 kPa)

CO

2

Simulated square wave

ETCO

2

40 mmHg (6.0 kPa)

IMCO

2

0 mmHg (0.0 kPa)

AWRR

25 rpm

Temperature

40°C (104°F)

No wave

Module

Test Numeric

Test Waveform

Monitor Setup

Test Signals

Pts

Refernce

Signals