Alarm Display

5-6 Alarm Functions

Ala

rm Func

tions

The alarm reminder clock starts counting from the beginning again
every time the

key is pressed. If an alarm has been

acknowledged by pressing the

key, and during the

next 3 minutes a new alarm (of whatever priority) occurs and is
silenced, the alarm reminder will sound 3 minutes later for the
highest priority alarm. The alarm reminder always functions for the
highest priority acknowledged alarm. The alarm reminder will not
sound if there is an active INOP or alarm present. For maximum
safety when monitoring patients, you must resolve active INOPs and
alarms quickly so that the alarm reminder function can operate.

Even when the alarms are suspended or individually switched off, the
corresponding parameter numeric will blink if an alarm condition
occurs and “Maintain Visual Alarms” is set to “Yes” in a special
Configuration Mode by your biomedical engineering department.

Note—

The central stations 78560A, M2350A and M2360A can be

configured to silence and reset the alarms of the bedside monitors. If
your system is connected to such a central station you can silence an
alarm at the central station. The central stations 78504A and 78508A can
silence and reset the alarms of the bedside monitor only if configured to
do so by a Philips service engineer or hospital biomedical engineer. This
is then permanent and cannot be altered by the user.

Re-alarm

2 or 3 mins

Full alarm tone
Alarm lamp
New Alarm Recording
Nurse call relay signals
A full alarm tone is reannounced after 2
or 3 minutes. The alarm tone will sound
continuously until it is silenced by
pressing the

key.

Alarm

Reminder

Reminder

Time

Alarm Behavior

Silence/Reset

Silence/Reset

Silence/Reset