Philips V24CT User Manual
Page 150

Using Philips Patient Care System with an Arrhythmia Computer
4-12 Other Patients
Ot
h
e
r Pa
ti
e
n
ts
Arrhythmia Alarms on the 78560A Arrhythmia Computer
Arrhythmia Alarms on the 78720 Arrhythmia Computer
Alarm
Minimum Condition Required for Alarm
RUN VPBs > X (**)
Run of more than X VPB's, where X is
adjustable from 2 to 9 using the Change
Limits display.
PACER NON-FUNCT
(**)
Missed beat without pace pulse in paced
patient
VPBs > XX/min (**)
More than XX VPBs per minute, where XX is
adjustable from 1 to 30 using the Change
Limit display.
Alarm
Minimum Condition Requiring an Alarm
EXTREME TACHY
Heart-rate >Tachy limit
RUN VPBs >9 (**)
Run of VPB's > 9 and heart-rate
RUN VPB's 3-9 (**)
Run of 3-9 VPB's
PACER NOT PACE (**)
Missed beat without pace pulse in paced
patient
VPB's > 23/min (**)
More than 23 VPB's/minute
VPB's > 11/min (**)
More than 11 VPB's/minute
VPB's > 5/min (**)
More than 5 VPB's/minute